Purifier H Synthetic High Flux Dialysers

  • Advanced Polyethersulfone membrane delivers excellent clearances
  • Excellent biocompatibility of advanced synthetic membrane
  • Oxygen-free gamma sterilization maintains membrane intehrit
  • Caps on dialysate ports help prevent comtamination
  • Wide rage of sizes to better matc performance to patient needs

Technical specifications & Performances

Clearances in vitro (mL / min)
Urea
Creatinine
Phosphate
Vitamin B12
Inulin
Qb(ml/min)
300
300
300
300
300
Qd(ml/min)
500
500
500
500
500
PURIFIER H 120
252
220
210
140
90
PURIFIER H 140
260
230
221
151
102
PURIFIER H 160
267
240
230
161
117
PURIFIER H 180
273
248
238
170
122
PURIFIER H 200
279
256
245
177
131
PURIFIER H 120 PURIFIER H 140 PURIFIER H 160 PURIFIER H 180 PURIFIER H 200
Effective Membrane area (m2)
1,2 m2
1,24 m2
1,6 m2
1,8 m2
2,0 m2
Priming Volume (mL)
67
74
87
96
108
UF Coefficient (mL/h/mmHg)
52
58
67
75
84
KoA Urea
(Qb 300 ml/min, Qd 500 ml/min.)
849
961
1083
1214
1382
Membrane
Polyethersulfone (PES)
Sterilisation
Gamma Rays
Max. TMP
600 mHg
Wall Thickness
30 µm
Inner Diameter
200 µm
  • Specifications & performance data at clearance in vitro (ml/min) Q =300m/min, Q = 500 ml/min, Q = 0 ml/min,T=37 oC.
  • KUF measurement using bovine / Human blood (Hc % 32;Protein 60g/L)
  • Performance data were measured in vitro according to standard ISO 8637

Dialysers Data Sheet

You can reach the catalog of dialysers products from the link below. To get an offer, you can contact us with the Offer Form.

Request a Quote